beautypg.com

Accu-Chek insulin pump User Manual

Page 164

background image

1. Press

f twice to turn off and to confirm the error.

2. Remove or disconnect the infusion set from your infusion site.

3. Prime the infusion set.

4. Reconnect the infusion set (tubing) at your infusion site.

5. Put your pump in RUN mode.

See the sections “Changing the Cartridge and Infusion Set” (page 73) and
“Priming the Infusion Set” (page 53) for more information.

Error E12: DATA INTERRUPTED

Data transfer between your pump and a PC was interrupted before it was
completed.

1. Press

f twice to turn off and to confirm the error.

2. Restart the data transfer.

Before you can put your pump in RUN mode, the data transfer must be successful.

!

If data transfer between compatible software and your
insulin pump is disrupted in any way, the configuration
may be incomplete and an error E12: DATA INTERRUPTED
occurs. The data transfer must be completed successfully
before you can put your insulin pump into RUN mode.

Chapter 7: Alerts and Errors

ACCU-CHEK Spirit Insulin Pump

163

CAUTION