Accu-Chek insulin pump User Manual

Page 173

background image

Appendixes

ACCU-CHEK Spirit Insulin Pump

172

IEC 60601

Compliance

Electromagnetic

Immunity Test Test Level

Level

Environment – Guidance

Radiated RF

3 V/m

10 V/m

Portable and mobile RF
communications equipment

IEC 61000-4-3

80 MHz to

80 MHz should be used no closer to

2.5 GHz

2.5 GHz

any part of the ACCU-CHEK
Spirit insulin pump, includ-

(IEC 60601-2-24) ing cables, than the recom-

mended separation distance
calculated from the equation
applicable to the frequency
of the transmitter.

Recommended separation
distance:

d

= 0.4 .

ǰҀ

P

80 MHz
to 800 MHz

d

= 0.7 .

ǰҀ

P

800 MHz
to 2.5 GHz

where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).

Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey (a), should be less
than the compliance level in
each frequency range (b).

Interface may occur in the
vicinity of equipment
marked with the following
symbol: