Accu-Chek insulin pump User Manual
Page 134
Lock Basal Rate Profiles
Your pump lets you change your hourly basal rates as needed. You may lock
your basal rate profiles to provide extra protection against accidentally changing
a basal rate during normal operation. When this function is activated, your hourly
basal rates cannot be changed.
Follow the steps below:
1. Press
d to move to the SETUP MENU ADVANCED screen.
Press
f to select.
2. Press
d to move to the BASAL RATE LOCK screen.
Press
a or s to turn the basal rate lock ON
or OFF.
3. Press
f to save and exit.
Chapter 5: User Menus
ACCU-CHEK Spirit Insulin Pump
133
Note:
If the basal rate lock is set to ON (
), the programming of the basal
rate profiles 1, 2, 3, 4, and 5 is blocked. To check if the basal rate lock
is ON, press
d to move to the PROGRAM BASAL RATE screen and
press
f to select. The
symbol in the BASAL RATE TOTAL screen
indicates the lock of the basal rate programming. (The
symbol appears
in the BASAL RATE TOTAL screen only.) Press
d to view the hourly
basal rates. The
a and s functions are blocked.