Quick reference table for errors – Accu-Chek insulin pump User Manual
Page 198
Index
ACCU-CHEK Spirit Insulin Pump
1
Stopping;
Insulin Delivery; 66
Priming; 55
System Check; 28-31
Temporary Basal Rate;
Cancelling; 97
History; 111
Over; 96, 153
Programming; 93-96
Time and Date; 32
American; 32, 130
European; 32, 130
Time and Date, Review; 150
Time Remaining; 112
Troubleshooting; 143-146
Tubing; 20, 132
UP Key; 9, 187
User Menu; 99-102
Advanced; 122
Custom; 136
Standard; 103
Water; 63
Contact with; 63
After Contact with; 64
Index
ACCU-CHEK Spirit Insulin Pump
197
200
Quick Reference Table for Errors
ACCU-CHEK Spirit Insulin Pump
Quick Reference Table for Errors
No. Error
What You Need to Do
E1
CARTRIDGE EMPTY
Change the cartridge.
E2
BATTERY DEPLETED
Change the battery.
E3
AUTOMATIC OFF
If necessary, put your ACCU-CHEK Spirit in
RUN mode.
E4
OCCLUSION
Immediately check your blood glucose level,
because insulin delivery has been interrupted.
If your blood glucose level is high, take
appropriate actions according to your doctor’s
or healthcare team’s instructions. See the
section “Error E4: OCCLUSION.”
E5
END OF OPERATION
Arrange for immediate replacement of your
insulin pump. Contact Pump
Support at
1-800-688-4578 to dis-
cuss your options for further use of the
ACCU-CHEK Spirit insulin pump. Contact
your healthcare professional for alternative
insulin therapy options.
E6
MECHANICAL ERROR
See the section “Error E6: MECHANICAL
ERROR.”
E7
ELECTRONIC ERROR
See the section “Error E7: ELECTRONIC
ERROR.”
E8
POWER INTERRUPT
See the section “Error E8: POWER INTER-
RUPT.”
E10
CARTRIDGE ERROR
See the section “Error E10: CARTRIDGE
ERROR.”
E11
SET NOT PRIMED
Prime your infusion set.
E12
DATA INTERRUPTED
Restart the data transfer.
E13
LANGUAGE ERROR
See the section “Error E13: LANGUAGE
ERROR.”