Accu-Chek insulin pump User Manual
Page 174
Accuracy of Flow Rate for Plastic Cartridges
Trumpet curve plotted from data after the end of the stabilization period
The trumpet curve shows the accuracy of the delivery rate in relation to the
observation period.
The measurements were made at a medium basal rate of 1.0 U/h according to
IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room
temperature.
The average deviation of the delivered amount (overall mean percentage flow
error) for U100 insulin is
≤
± 5%.
Appendixes
ACCU-CHEK Spirit Insulin Pump
173