beautypg.com

Accu-Chek insulin pump User Manual

Page 151

background image

Alert A2: BATTERY LOW

The battery power is low. Be prepared for a battery change as quickly as possible.
The battery low symbol (

) appears in the RUN or STOP screen as a reminder

until you change the battery.

1. Press

f twice to turn off and to confirm the alert.

Replace the battery as soon as possible (within a
few hours). Failure to change the battery within a
few hours of receiving the alert could result in
additional error messages.

2. If necessary, put your pump in RUN mode.

See the section “Accessories and Disposables – Battery” (page 21) for more
information on recommended batteries. See the section “Inserting (or Changing)
the Battery” (page 25) for instructions on how to change your battery.

Alert A3: REVIEW TIME AND DATE

The pump was without power for more than one hour. The ACCU-CHEK Spirit
insulin pump settings and history data are saved in memory regardless of
power loss. However, it may be necessary to enter the time and date again.

1. Press

f twice to turn off and to confirm the alert.

Review the time and date and correct them,
if they are wrong.

2. If necessary, put your pump in RUN mode.

See the section “Setting Time and Date” (page 32) for more information.

Chapter 7: Alerts and Errors

ACCU-CHEK Spirit Insulin Pump

150