Routing signals to counter n hw arm input, Counter n internal output and counter n tc signals, Frequency output signal – National Instruments DAQ M Series User Manual
Page 91: Routing frequency output to a terminal, Routing signals to counter n hw arm input -29, Frequency output signal -29, Routing frequency output to a terminal -29

Chapter 7
Counters
© National Instruments Corporation
7-29
waiting for the Gate signal when it is armed. Counter output operations can
use the arm signal in addition to a start trigger.
Software can arm a counter or configure counters to be armed on a
hardware signal. Software calls this hardware signal the Arm Start Trigger.
Internally, software routes the Arm Start Trigger to the Counter n HW Arm
input of the counter.
Routing Signals to Counter n HW Arm Input
Any of the following signals can be routed to the Counter n HW Arm input.
•
RTSI <0..7>
•
Input PFI <0..5>
•
ai/ReferenceTrigger
•
ai/StartTrigger
•
PXI_STAR
Counter 1 Internal Output can be routed to Counter 0 HW Arm. Counter 0
Internal Output can be routed to Counter 1 HW Arm.
Some of these options may not be available in some driver software.
Counter n Internal Output and Counter n TC Signals
The Counter n Internal Output signal changes in response to Counter n TC.
The two software-selectable output options are pulse output on TC and
toggle output on TC. The output polarity is software-selectable for both
options.
Routing Counter n Internal Output to an Output
Terminal
You can route Counter n Internal Output to any output PFI <6..9> or
RTSI <0..7> terminal. All output PFIs are set to high-impedance at startup.
Frequency Output Signal
The Frequency Output (FREQ OUT) signal is the output of the frequency
output generator.
Routing Frequency Output to a Terminal
You can route Frequency Output to any output PFI <6..9> terminal. All
PFIs are set to high-impedance at startup.