Figure 4-6. enabling multimode scanning in labview – National Instruments DAQ M Series User Manual
Page 31

Chapter 4
Analog Input
© National Instruments Corporation
4-7
Caution The maximum input voltage and current ratings of AI signals with respect to 
AI GND (and for differential signals with respect to each other) and earth/chassis ground 
are listed in the Maximum Working Voltage section of the NI 6236 Specifications. 
Exceeding the maximum input voltage or maximum working voltage of AI signals distorts 
the measurement results. Exceeding the maximum input voltage or maximum working 
voltage rating also can damage the device and the computer. Exceeding the maximum input 
voltage can cause injury and harm the user. NI is not liable for any damage or injuries 
resulting from such signal connections.
AI ground-reference setting is sometimes referred to as AI terminal 
configuration.
Configuring AI Ground-Reference Settings in Software
You can program channels on an M Series device to acquire with different 
ground references.
To enable multimode scanning in LabVIEW, use
NI-DAQmx Create
Virtual Channel.vi
of the NI-DAQmx API. You must use a new VI for
each channel or group of channels configured in a different input mode. In 
Figure 4-6, channel 0 is configured in differential mode and 
aignd_vs_aignd reads the internal analog input ground-reference of the 
device.
Figure 4-6. Enabling Multimode Scanning in LabVIEW
To configure the input mode of your current measurement using the DAQ 
Assistant, use the Terminal Configuration drop-down list. Refer to the 
DAQ Assistant Help for more information about the DAQ Assistant.
To configure the input mode of your current measurement using the 
NI-DAQmx C API, set the terminalConfig property. Refer to the 
NI-DAQmx C Reference Help for more information.