For example, suppose all channels in a system use a –20 to 20 mA input 
range. The signals on channels 0 and 2 vary between 18 and 19 mA. The 
signals on channels 1 and 3 vary between –18 and 0 mA. Scanning 
channels in the order 0, 2, 1, 3 will produce more accurate results than 
scanning channels in the order 0, 1, 2, 3.
Avoid Scanning Faster Than Necessary
Designing your system to scan at slower speeds gives the PGIA more time 
to settle to a more accurate level. Here are two examples to consider.
Example 1
Averaging many AI samples can increase the accuracy of the reading by 
decreasing noise effects. In general, the more points you average, the more 
accurate the final result will be. However, you may choose to decrease the 
number of points you average and slow down the scanning rate.
Suppose you want to sample 10 channels over a period of 40 ms and 
average the results. You could acquire 500 points from each channel at a 
scan rate of 125 kS/s. Another method would be to acquire 1,000 points 
from each channel at a scan rate of 250 kS/s. Both methods take the same 
amount of time. Doubling the number of samples averaged (from 500 to 
1,000) decreases the effect of noise by a factor of 1.4 (the square root of 2). 
However, doubling the number of samples (in this example) decreases the 
time the PGIA has to settle from 8 µs to 4 µs. In some cases, the slower scan 
rate system returns more accurate results.
Example 2
If the time relationship between channels is not critical, you can sample 
from the same channel multiple times and scan less frequently. For 
example, suppose an application requires averaging 100 points from 
channel 0 and averaging 100 points from channel 1. You could alternate 
reading between channels—that is, read one point from channel 0, then one 
point from channel 1, and so on. You also could read all 100 points from 
channel 0 then read 100 points from channel 1. The second method 
switches between channels much less often and is affected much less by 