8 tempconf configuration - pv3 – Honeywell SMV 3000 User Manual
Page 78

SMV 3000 Transmitter User’s Manual
1/99
66
6.8
TempConf Configuration - PV3,
Continued
ATTENTION
•
For a reverse range, enter the upper range value as the LRV and the
lower range value as the URV. For example, to make a 0 to 500 °F
range a reverse range, enter 500 as the LRV and 0 as the URV.
•
The URV changes automatically to compensate for any changes in the
LRV and maintain the present span (URV – LRV). See Figure 23 for
an example.
•
If you must change both the LRV and URV, always change the LRV
first. However, if the change in the LRV would cause the URV to
exceed the URL, you would have to change the URV to narrow the
span before you could change the LRV
Figure 23
Example of LRV and URV Interaction
Current Range Settings
LRL
LRV
SPAN
URV
URL
-328
-100
257
600
842 oF
Range Settings After LRV is Changed to Zero (0)
LRL
LRV
SPAN
URV URL
-328
-100 0
257
600 700 842 oF
Continued on next page