Source 57, Source, (trigger:source), 57, Source – HP E1367A User Manual
Page 57

:SOURce
TRIGger:SOURce BUS | EXTernal | HOLD | IMMediate
specifies the
trigger source to advance the channel list during scanning.
Parameters
Parameter Name
Parameter Type
Range of Values
BUS
discrete
*TRG or GET command
EXTernal
discrete
“Event In” BNC port
HOLD
discrete
Hold triggering
IMMediate
discrete
Immediate triggering
Comments
•
Enabling the Trigger Source: The
TRIGger:SOURce
command
only selects the trigger source. The
INITiate[:IMMediate]
command
enables the trigger source.
•
Using the
TRIG
Command: You can use
TRIGger[:IMMediate]
to
advance the scan when
TRIGger:SOURce HOLD
or
TRIGger:SOURce BUS
is selected.
•
Using External Trigger Inputs: With
TRIGger:SOURce EXTernal
selected, only one switchbox at a time can use the external trigger
input at the HP E1300A/E1301A mainframe “ Event In” BNC port.
The trigger input is assigned to the first switchbox that requested the
external trigger source (with a
TRIGger:SOURce EXTernal
command).
•
Assigning External Trigger: A switchbox assigned with
TRIGger:SOURce EXTernal
remains assigned to that source until the
switchbox trigger source is changed to
BUS
,
HOLD
, or
IMMediate
.
When the source is changed, the external trigger source is available
to the next switchbox which requests it (with a
TRIGger:SOURce EXTernal
command). If a switchbox requests an
external trigger input already assigned to another switchbox an error
is generated.
•
Using Bus Triggers: To trigger the switchbox with
TRIGger:SOURce BUS
selected, use the IEEE 488.2 common
command
*TRG
or the HP-IB Group Execute Trigger (
GET
)
command.
•
Trig Out Port Shared by Switchboxes: See the
OUTPut
command.
•
Related Commands:
ABORt
,
[ROUTe:]SCAN
,
TRIGger
•
*RST Condition:
TRIGger:SOURce IMMediate
Chapter 5
HP E1366A/E1367A RF Multiplexers Command Reference 57