beautypg.com

Trigger 56, Immediate] 56, Trigger subsystem, 56 - 58 – HP E1367A User Manual

Page 56: Trigger[:immediate], 56, 60, Immediate, (trigger[:immediate]), 56, Trigger, Immediate

background image

TRIGger

The

TRIGger

subsystem controls the triggering operation of RF multiplexer

modules in a switchbox.

Subsystem Syntax

TRIGger

[:IMMediate]
:SOURce BUS | EXTernal | HOLD | IMMediate
:SOURce?

[:IMMediate]

TRIGger[:IMMediate]

causes a trigger to occur when the defined trigger

source is

TRIGger:SOURce HOLD

or

TRIGger:SOURce BUS

.

Comments

Executing the

TRIGger[:IMMediate]

Command: A channel list

must be defined with

[ROUTe:]SCAN <

channel_list

>

and an

INITiate[:IMMediate]

command must be executed before

TRIGger[:IMMediate]

will be executed.

HOLD

or

BUS

Source Remains: The

TRIGger:SOURce HOLD

or

TRIGger:SOURce BUS

commands remain in effect, if selected, after

triggering a switchbox with the

TRIGger[:IMMediate]

command.

Related Commands:

INITiate

,

[ROUTe:]SCA

N

Example

Advancing Scan Using the

TRIGger

Command

This example uses the

TRIGger

command to advance the scan of a

single-module switchbox from channel 00 through 03. Since

TRIGger:SOURce HOLD

is set, the scan is advanced one channel each time

TRIGger

is executed.

TRIG:SOUR HOLD

! Sets trigger source to

HOLD

SCAN (@100:103)

! Defines channel list

INIT

! Begin scan, close channel 00

loop statement

! Start count loop

TRIG

! Advance scan to next channel

increment loop

! Increment loop count

56 HP E1366A/E1367A RF Multiplexers Command Reference

Chapter 5