Display 43, Monitor[:state] 43, Display subsystem, 43 - 44 – HP E1367A User Manual
Page 43: Display:monitor[:state], 43, 60, Display, 43, Monitor[:state], (display:monitor[:state]), 43, Display, Monitor[:state

DISPlay
The
DISPlay
subsystem monitors the channel state of a selected module
(or card) in a switchbox. This subsystem operates only with mainframes
which have a display, such as the HP 75000 Series B Mainframe
(HP E1301A).
Subsystem Syntax
DISPlay
:MONitor
[:STATe] <
boolean
>
:CARD <
number
> | AUTO
:MONitor[:STATe]
DISPlay:MONitor[:STATe] <
boolean
>
turns the monitor mode on or off.
Parameters
Parameter Name
Parameter Type
Range of Values
ON | OFF | 1 | 0
boolean
ON | OFF | 1 | 0
Comments
•
Monitoring Switchbox Channels:
DISPlay[:STATe] ON
or
DISPlay[:STATe] 1
turns the monitor mode on to show the channel
state of the selected module.
DISPlay[:STATe] OFF
or
DISPlay[:STATe] 0
turns the monitor mode off.
•
Selecting the Module to be Monitored: Use the
DISPlay:MONitor:CARD <
number
> AUTO
command to select the
module.
•
Monitor Mode on an HP E1301A Mainframe Display: A typical
display for the HP E1366A or HP E1367A RF Multiplexer module
follows, where channel 00 (in bank 0) and channel 13 (in bank 1) are
closed.
B0:0 B1:3
•
*RST Condition:
DISPlay:MONitor[:STATe] OFF | 0
Example
Enabling the Monitor Mode
DISP:MON:CARD 2
! Selects module #2 in a switchbox
DISP:MON 1
! Turns the monitor mode ON
Chapter 5
HP E1366A/E1367A RF Multiplexers Command Reference 43