beautypg.com

Continuous? 46, Immediate] 46, Initiate:continuous?, 46, 60 – HP E1367A User Manual

Page 46: Initiate[:immediate], 46, 60, Continuous?, (initiate:continuous?), 46, Immediate, (initiate[:immediate]), 46, Continuous, Immediate

background image

:CONTinuous?

INITiate:CONTinuous?

queries the scanning state. With continuous

scanning enabled, the command returns 1. With continuous scanning
disabled, the command returns 0.

Example

Query Continuous Scanning State

This example enables continuous scanning of a switchbox and queries the
state. Since continuous scanning is enabled,

INIT:CONT?

returns 1.

INIT:CONT ON

! Enable continuous scanning

INIT:CONT?

! Query continuous scanning state

[:IMMediate]

INITiate[:IMMediate]

starts the scanning cycle and closes the first channel

in the channel list. Successive triggers from the source selected by the

TRIGger:SOURce

command advances the scan through the channel list.

Comments

Starting the Scanning Cycle: The

INITiate[:IMMediate]

command

starts scanning by closing the first channel in the channel list. Each
trigger received advances the scan to the next channel in the channel
list. An invalid channel list definition generates an error (see

[ROUTe:]SCAN

command).

Stopping Scanning Cycles: See the

ABORt

command.

Example

Enabling a Single Scan

This example enables a single scan of channels 00 through 03 of a
single-module switchbox. The trigger source to advance the scan is
immediate (internal) triggering set with (default)

TRIGger:SOURce

IMMediate.

SCAN (@100:103)

! Scan channels 00-03

INIT

! Begin scan, close channel 00 (use
immediate triggering)

46 HP E1366A/E1367A RF Multiplexers Command Reference

Chapter 5

This manual is related to the following products: