Initiate 45, Continuous 45, Initiate subsystem, 45 - 46 – HP E1367A User Manual
Page 45: Initiate:continuous, 45, 60, Continuous, (initiate:continuous), 45, Initiate, Continuous

INITiate
The
INITiate
subsystem selects continuous scanning cycles and starts the
scanning cycle.
Subsystem Syntax
INITiate
:CONTinuous ON | OFF | 1 | 0
:CONTinuous?
[:IMMediate]
:CONTinuous
INITiate:CONTinuous ON | OFF | 1 | 0
enables or disables continuous
scanning cycles for the switchbox.
Parameters
Parameter Name
Parameter Type
Range of Values
ON | OFF | 1 | 0
boolean
ON | OFF | 1 | 0
Comments
•
Continuous Scanning Operation: Continuous scanning is enabled
with the
INITiate:CONTinuous ON
or
INITiate:CONTinuous 1
command. Sending the
INITiate[:IMMediate]
command closes the
first channel in the channel list.
•
Each trigger from a trigger source selected by the
TRIGger:SOURce
command advances the scan through the channel list. A trigger at
the end of the channel list closes the first channel in the channel list
and the scan cycle repeats.
•
Non-Continuous Scanning Operation: This is enabled with the
INITiate:CONTinuous OFF
or
INIT:CONT 0
command. Sending the
INIT:IMM
command closes the first channel in the channel list.
Each trigger from a trigger source selected by the
TRIGger:SOURce
command advances the scan through the channel list. At the end of
the scanning cycle, all channels in the channel list are closed.
•
Stopping Continuous Scans: See the
ABORt
command.
•
Related Commands:
ABORt
,
ARM:COUNt
,
TRIGger
•
*RST Condition:
INITiate:CONTinuous OFF
Example
Enabling Continuous Scans
This example enables continuous scanning of channels 00 through 03 of a
single-module switchbox. Since
TRIGger:SOURce IMMediate
(default) is
set, use an interface clear command (such as
CLEAR
) to stop the scan.
INIT:CONT ON
! Enable continuous scanning
SCAN (@100:103)
! Define channel list
INIT
! Start scan cycle, close channel 00
Chapter 5
HP E1366A/E1367A RF Multiplexers Command Reference 45