SCPI Command Reference

This section describes the Standard Commands for Programmable
Instruments (SCPI) commands for the RF multiplexers. Commands are
listed alphabetically by subsystem and within each subsystem.

ABORt

The

ABORt

subsystem stops a scan in progress when the scan is enabled via

the interface, and the trigger modes are

TRIGger:SOURce BUS

or

TRIGger:SOURce HOLD

.

Subsystem Syntax

ABORt

Comments

ABORt

operation invalidates the current channel list and sets

ARM:COUNt 1

(one scanning cycle per

INITiate

command), sets

INITiate CONTinuous OFF

(no continuous scanning cycles), and sets

TRIGger:SOURce IMMediate

(continuous internal triggering).

Stopping Scans Enabled from Interface: When a scan is enabled
from the interface, use an interface

CLEAR

command or the

HP E1301A front panel “ Reset Instr” or “Clear Instr” key to stop
the scan.

When the scan is enabled from the interface and the trigger source is

TRIGger:SOURce BUS

or

TRIGger:SOURce HOLD

, use

ABORt

or

the HP E1301A front panel “ Reset Instr” or “Clear Instr” key to
stop the scan.

Stopping Scans Enabled from Front Panel: When a scan is
enabled from the HP E1301A front panel, execute

*RST

entered via

the interface or the front panel “ Reset Instr” or “ Clear Instr” key to
stop the scan.

Related Commands:

ARM

,

INITiate:CONTinuous

,

[ROUTe:]SCAN

,

TRIGger

Example

Stopping a Scan with

ABORt

TRIG:SOUR BUS

!

*TRG

command is trigger source

INIT:CONT ON

! Set continuous scanning

SCAN (@100:103)

! Scan channels 00-03

INIT

! Start scan, close channel 00

.

.

ABOR

! Abort scan in progress

40 HP E1366A/E1367A RF Multiplexers Command Reference

Chapter 5