Page 32

Sets Number of Scanning Cycles
ARM:COUN<
number
>
sets 1 to 32767 scans
through the channel list per
INIT
command.
Default is one scanning cycle.
Selects the Trigger Source
TRIG:SOUR
defines the trigger source to
advance the scan. Default is
TRIG:SOUR IMM.
TRIG:SOUR?
queries the current trigger source.
Sources are:
BUS = *TRG
via HP-IB interface
EXT =
Event In connector
HOLD =
Hold triggering
IMM =
Automatic advance
Selects Continuous Scanning Cycles
Use
INIT:CONT ON
or
INIT:CONT 1
to enable
continuous scanning cycles. Default made is
fixed number of scans, with number set by
ARM:COUN<
number
>.
Selects Scan Mode
SCAN:MODE
sets the RF multiplexers for
NONE
,
VOLT
,
RES
, or
FRES
.
NONE
,
VOLT
,
and
RES
have no effect on RF multiplexer
operation.
FRES
sets "paired-channel"
scanning.
Selects the Scan List
SCAN<
channel_list
>
defines channels to
be scanned using the trigger source set by
the
TRIG:SOUR
command.
Enables Scanning
INIT
starts scanning for any trigger source
and closes the first channel in the channel list
.
Advances the Scan
TRIG
can be used to advance the scan for
TRIG:SOUR BUS
or
TRIG:SOUR HOLD.
The
command is not functional when
TRIG:SOUR IMM
or
TRIG:SOUR EXT
is set.
Figure 4-1. Scanning Channels Commands
32 Understanding the HP E1366A/E1367A RF Multiplexers