10
Specifications
Topics covered in this chapter:
•
•
•
•
•
10.1
Electrical connections
Electrical connections
Table 10-1:
Type
Description
Input/output connections
Three pairs of wiring terminals for transmitter outputs. Screw
connectors accept one or two solid conductors, 14 to 12 AWG
(2.5 to 4.0 mm
2
); or one or two stranded conductors, 22 to
14 AWG (0.34 to 2.5 mm
2
).
Power connections
One pair of wiring terminals accepts AC or DC power.
One internal ground lug for power-supply ground wiring.
Screw connectors accept one or two solid conductors, 14 to
12 AWG (2.5 to 4.0 mm
2
); or one or two stranded conductors, 22
to 14 AWG (0.34 to 2.5 mm
2
).
Digital communications main-
tenance connections
Two clips for temporary connection to the service port.
Core processor connection
The transmitter has two pairs of terminals for the 4-wire connec-
tion to the core processor:
•
One pair is used for the RS-485 connection to the core pro-
cessor
•
One pair is used to supply power to the core processor
Plug terminals accept solid or stranded conductors, 24 to
12 AWG (0.40 to 3.5 mm
2
).
Power
Self-switching AC/DC input, automatically recognizes supply
voltage
•
85 to 265 VAC, 50/60 Hz, 6 watts typical, 11 watts maximum
•
18 to 100 VDC, 6 watts typical, 11 watts maximum
•
Complies with low voltage directive 2006/95/EC per EN
61010-1 (IEC 61010-1) with amendment 2, and Installation
(Overvoltage) Category II, Pollution Degree 2
Specifications
Installation Manual
109