Environmental limits, Hazardous area classifications, Power requirements – Emerson MICRO MOTION 1700 User Manual
Page 11: 3 environmental limits, 4 hazardous area classifications, 5 power requirements

Output options for Model 1500 and Model 2500 transmitters
Table 1-4:
Letter
Description
A
Analog outputs – one mA, one frequency, one RS-485
B
Configurable I/O channels (default configuration of two mA, one frequency)
C
Configurable I/O channels (custom configuration )
1.3
Environmental limits
Environmental specifications
Table 1-5:
Type
Value
Ambient temperature limits
–40 to +140 °F (–40 to +60 °C)
Humidity limits
5 to 95% relative humidity, non-condensing at 140 °F (60 °C)
Vibration limits
Meets IEC68.2.6, endurance sweep, 5 to 2000 Hz, 50 sweep cy-
cles at 1.0 g
EMI effects
Complies with EMC Directive 2004/108/EC per EN 61326 Indus-
trial
Complies with NAMUR NE-21 (22.08.2007)
Ambient temperature effect on
analog outputs
On mA output: ±0.005% of span per °C
If possible, install the transmitter in a location that will prevent direct exposure to sunlight.
The environmental limits for the transmitter may be further restricted by hazardous area
approvals.
1.4
Hazardous area classifications
If you plan to mount the transmitter in a hazardous area:
•
Verify that the transmitter has the appropriate hazardous area approval. Each
transmitter has a hazardous area approval tag attached to the transmitter housing.
•
Ensure that any cable used between the transmitter and the sensor meets the
hazardous area requirements.
1.5
Power requirements
Self-switching AC/DC input, automatically recognizes supply voltage
•
85 to 265 VAC, 50/60 Hz, 6 watts typical, 11 watts maximum
Planning
Installation Manual
7