Environmental limits, Physical specifications, 4 environmental limits – Emerson MICRO MOTION 1700 User Manual
Page 117: 5 physical specifications

10.4
Environmental limits
Environmental specifications
Table 10-9:
Type
Value
Ambient temperature limits
–40 to +140 °F (–40 to +60 °C)
Humidity limits
5 to 95% relative humidity, non-condensing at 140 °F (60 °C)
Vibration limits
Meets IEC68.2.6, endurance sweep, 5 to 2000 Hz, 50 sweep cy-
cles at 1.0 g
EMI effects
Complies with EMC Directive 2004/108/EC per EN 61326 Indus-
trial
Complies with NAMUR NE-21 (22.08.2007)
Ambient temperature effect on
analog outputs
On mA output: ±0.005% of span per °C
If possible, install the transmitter in a location that will prevent direct exposure to sunlight.
The environmental limits for the transmitter may be further restricted by hazardous area
approvals.
10.5
Physical specifications
Physical specifications
Table 10-10:
Type
Description
Mounting options
Field-mount
•
Integrally mounted to a Micro Motion F-Series or R-Series
sensor
•
Remotely mounted to any 4-wire or 9-wire Micro Motion
Coriolis sensor
Housing
NEMA 4X (IP66) polyurethane-painted cast aluminum
Weight (4-wire remote-mount
option)
8 lb (3.6 kg)
Weight (9-wire remote-mount
option)
14 lb (6.3 kg)
Cable gland entrances
1/2˝–14 NPT or M20
× 1.5 female conduit ports for outputs and
power supply
3/4˝–14 NPT female conduit port for sensor/core processor ca-
ble
Specifications
Installation Manual