Emerson MICRO MOTION 1700 User Manual
Page 94

Safe area mA output load resistance values
Figure 8-9:
1000
900
800
700
600
500
400
300
200
100
0
12
14
16
18
20
22
24
26
28
30
OPERATING REGION
Supply voltage VDC (Volts)
External resistor R
load
(Ohms)
R
max
= (V
supply
– 12)/0.023
Min. 250W and 17.5V required for HART communication
I/O wiring for Model 1700 and Model 2700 transmitters with intrinsically safe outputs