ETA Systems FC01-LQ User Manual
Page 55

Flow Meter FC01-LQ
Flow Meter FC01-LQ
Technical data
49
Technical data
49
Inrush current:
I
p
= typ. 3 A (20 µs)
Switch-off current:
I
kipp
= typ. 0.75 A
Rated power consumption:
P
n
= 4.1 W with zero flow, voltage outputs
P
n
= 4.8 W max. flow (end of measuring range),
voltage outputs
Insulation voltage:
supply input - central electronic unit
≥ DC 500 V
8.3 Analogue outputs
The analogue outputs are physically isolated from each other as well as from the electronic
control unit FC01-LQ.
Pin selection for analogue outputs V1, V2 and C1
Signal name
Pin XAO
NC
1
analogue output 1 - flow rate
2
reference ground 1
3
shield 1 *
4
shield 2 *
5
analogue output 2 - temperature
6
reference ground 2
7
NC
8
NC - not used
Analogue output 1 - ANA OUT FLOW (flow output)
Analogue output 2 - ANA OUT TEMP. (temperature output)
* Shield ungrounded - apply on one side only.
The output is reverse polarity protected.
Insulation voltage:
analogue output - analogue output DC 500 V
analogue output - central electronic unit DC 500 V
Inrush current:
I
p
= typ. 3 A (20 µs)
Switch-off current:
I
kipp
= typ. 0.75 A
Rated power consumption:
P
n
= 4.1 W with zero flow, voltage outputs
P
n
= 4.8 W max. flow (end of measuring range),
voltage outputs
Insulation voltage:
supply input - central electronic unit
≥ DC 500 V
8.3 Analogue outputs
The analogue outputs are physically isolated from each other as well as from the electronic
control unit FC01-LQ.
Pin selection for analogue outputs V1, V2 and C1
Signal name
Pin XAO
NC
1
analogue output 1 - flow rate
2
reference ground 1
3
shield 1 *
4
shield 2 *
5
analogue output 2 - temperature
6
reference ground 2
7
NC
8
NC - not used
Analogue output 1 - ANA OUT FLOW (flow output)
Analogue output 2 - ANA OUT TEMP. (temperature output)
* Shield ungrounded - apply on one side only.
The output is reverse polarity protected.
Insulation voltage:
analogue output - analogue output DC 500 V
analogue output - central electronic unit DC 500 V