ETA Systems FC01-LQ User Manual
Page 24

Flow Meter FC01-LQ
Flow Meter FC01-LQ
18
Installation
18
Installation
XAO - - Analogue outputs
Connection by 8 pole connector; max. 1.5 mm
2
; LiYCY 2 x 0.25 mm
2
cable recommended
for each analogue output
Pin selection for analogue outputs (option: V1, V2, C1)
Pin No.
Signal name
Function
1
NC
none
2
ANAO1
analogue output 1 - flow
3
ANA1GND
reference potential for analogue output 1
4
SGNDA1
shield ground for analogue output 1 (ungrounded) *
5
SGNDA2
shield ground for analogue output 2 (ungrounded) *
6
ANAO2
analogue output 2 - temperature
7
ANA2GND
reference potential for analogue output 2
8
NC
none
* Apply shield on one side only.
XAH - Limit switch signal outputs - relay outputs - change over contacts
Connection by 8 pole connector; max. 1.5 mm
2
; LiYCY 3 x 0.38 mm
2
cable recommended
for each signal output
Pin No.
Signal name
Function
1
SGNDL1
shield ground 1
2
LIM1
non-inverted signal output 1 (N/O)
3
LIM1COM
common change over input 1
4
/LIM1
inverted signal output 1 (N/C)
5
SGNDL2
shield ground 2
6
LIM2
non-inverted signal output 2 (N/O)
7
LIM2COM
common change over input 2
8
/LIM2
inverted signal output 2 (N/C)
XAH - Limit switch signal outputs - transistor outputs NPN, freely connectable as emitter (-)
and collector (+) have been brought out separately.
Connection by 8 pole connector; max. 1.5 mm
2
; LifYCY 4 x 2 x 0.2 mm
2
cable recommended
Pin No.
Signal name
Function
1
/ERROR E
summarized error indication - emitter terminal
2
/ERROR C
summarized error indication - collector terminal
3
/BUSY/PULSE E
availability signal or frequency output -
emitter terminal
4
/BUSY/PULSE C
availability signal or frequency output -
collector terminal
XAO - Analogue outputs
Connection by 8 pole connector; max. 1.5 mm
2
; LiYCY 2 x 0.25 mm
2
cable recommended
for each analogue output
Pin selection for analogue outputs (option: V1, V2, C1)
Pin No.
Signal name
Function
1
NC
none
2
ANAO1
analogue output 1 - flow
3
ANA1GND
reference potential for analogue output 1
4
SGNDA1
shield ground for analogue output 1 (ungrounded) *
5
SGNDA2
shield ground for analogue output 2 (ungrounded) *
6
ANAO2
analogue output 2 - temperature
7
ANA2GND
reference potential for analogue output 2
8
NC
none
* Apply shield on one side only.
XAH - Limit switch signal outputs - relay outputs - change over contacts
Connection by 8 pole connector; max. 1.5 mm
2
; LiYCY 3 x 0.38 mm
2
cable recommended
for each signal output
Pin No.
Signal name
Function
1
SGNDL1
shield ground 1
2
LIM1
non-inverted signal output 1 (N/O)
3
LIM1COM
common change over input 1
4
/LIM1
inverted signal output 1 (N/C)
5
SGNDL2
shield ground 2
6
LIM2
non-inverted signal output 2 (N/O)
7
LIM2COM
common change over input 2
8
/LIM2
inverted signal output 2 (N/C)
XAH - Limit switch signal outputs - transistor outputs NPN, freely connectable as emitter (-)
and collector (+) have been brought out separately.
Connection by 8 pole connector; max. 1.5 mm
2
; LifYCY 4 x 2 x 0.2 mm
2
cable recommended
Pin No.
Signal name
Function
1
/ERROR E
summarized error indication - emitter terminal
2
/ERROR C
summarized error indication - collector terminal
3
/BUSY/PULSE E
availability signal or frequency output -
emitter terminal
4
/BUSY/PULSE C
availability signal or frequency output -
collector terminal