7.2 Potential errors

Independent of the priority group, all errors found are indicated with their relevant number.

In order to facilitate operation, the last error is stored in a non-volatile memory. The stored error
may be retrieved and deleted in the main menu.

If a combination of errors occurs simultaneously, they are indicated or stored in the error memory
observing the following sequence.

Priority group I

Error

Cause

Rectification

No. 1

No system parameter available

Return to supplier.

No. 2

Incorrect test sum of parameter

Return to supplier.

memory

No. 3

Incorrect test sum of program

Return to supplier.

memory

No. 4

Incorrect test sum of data memory

Return to supplier.

No. 5

Internal controller error

Return to supplier.

Priority group II

Error

Cause

Rectification

No. 10

Sensor not connected; or cable

Check cable or replace sensor.

between FC01-LQ and sensor
defective; or defective sensor

Sensor selected (configuration)

Correct sensor selection in

differs from sensor connected

configuration menu

No. 21

Medium temperature too high

7.2 Potential errors

Independent of the priority group, all errors found are indicated with their relevant number.

In order to facilitate operation, the last error is stored in a non-volatile memory. The stored error
may be retrieved and deleted in the main menu.

If a combination of errors occurs simultaneously, they are indicated or stored in the error memory
observing the following sequence.

Priority group I

Error

Cause

Rectification

No. 1

No system parameter available

Return to supplier.

No. 2

Incorrect test sum of parameter

Return to supplier.

memory

No. 3

Incorrect test sum of program

Return to supplier.

memory

No. 4

Incorrect test sum of data memory

Return to supplier.

No. 5

Internal controller error

Return to supplier.

Priority group II

Error

Cause

Rectification

No. 10

Sensor not connected; or cable

Check cable or replace sensor.

between FC01-LQ and sensor
defective; or defective sensor

Sensor selected (configuration)

Correct sensor selection in

differs from sensor connected

configuration menu

No. 21

Medium temperature too high