beautypg.com

2 connecting the load (local sense), 3 connecting the load (remote sense), Connecting the load (local sense) – KEPCO RCW 350W Series User Manual

Page 9: Connecting the load (remote sense)

background image

RCW350W 120607

7

3.2 CONNECTING THE LOAD (LOCAL SENSE)

To connect the load for local sensing, the shorting links must be maintained between the (+) S and
(+) M terminals, and between the (–) S and (–) M terminals. The load is connected across the DC
output (+) and (–) terminals.

3.3 CONNECTING THE LOAD (REMOTE SENSE)

The load is connected as shown in Figure 5. Error sensing may be done at the load terminals to
compensate for voltage loss in the connecting wires. The shorting links must be removed from the
(±) S Sense terminals.

FIGURE 5 CONNECTIONS FOR REMOTE SENSING WITH THE RCW 350W POWER SUPPLY