beautypg.com

Specifications, Electrical characteristics – RIGOL DS6000 Series User Manual

Page 39

background image

RIGOL

RP1001C User’s Guide

9

Specifications


These characteristics apply to an adjusted RP1001C current probe
installed on an oscilloscope of any brand. The oscilloscope must be
warmed up for at least 20 minutes and be in an environment with the
temperature at 10℃-30℃ and the humidity at 0-80%.

Electrical Characteristics

Current Range

10mV/A, 100mV/A

DC Accuracy, typical

100mV/A: ±3% ±50mA
(50mA to 10A peak range)
10mV/A: ±4% ±50mA
(500mA to 40A peak range)
100mV/A: ±15% maximum
(40A peak to 100A peak range)

Gain versus
frequency, typical

See Appendix 1 Gain versus Frequency at
1 A Peak (Typical)

Max. Working Current

See Voltage and Current Ratings

Max. Working Voltage

300V CAT III, 600V CAT II, see Voltage and
Current Ratings

Max. Floating Voltage

See Voltage and Current Ratings

Frequency Range

DC to 300kHz (-3dB)

Battery Type and Life,
typical

9V NEDA 1604A, IEC 6LR61
13 hours (1 each)

DC Signal Linearity,
typical

See Appendix 3 DC Signal Linearity in
10mV/A Range (Typical)

Phase versus
frequency, typical

See Appendix 4 Phase versus Frequency
at 1 A Peak (Typical)