Specifications, Electrical characteristics – RIGOL DS6000 Series User Manual
Page 39

RIGOL
RP1001C User’s Guide
9
Specifications
These characteristics apply to an adjusted RP1001C current probe
installed on an oscilloscope of any brand. The oscilloscope must be
warmed up for at least 20 minutes and be in an environment with the
temperature at 10℃-30℃ and the humidity at 0-80%.
Electrical Characteristics
Current Range
10mV/A, 100mV/A
DC Accuracy, typical
100mV/A: ±3% ±50mA
(50mA to 10A peak range)
10mV/A: ±4% ±50mA
(500mA to 40A peak range)
100mV/A: ±15% maximum
(40A peak to 100A peak range)
Gain versus
frequency, typical
See Appendix 1 Gain versus Frequency at
1 A Peak (Typical)
Max. Working Current
See Voltage and Current Ratings
Max. Working Voltage
300V CAT III, 600V CAT II, see Voltage and
Current Ratings
Max. Floating Voltage
See Voltage and Current Ratings
Frequency Range
DC to 300kHz (-3dB)
Battery Type and Life,
typical
9V NEDA 1604A, IEC 6LR61
13 hours (1 each)
DC Signal Linearity,
typical
See Appendix 3 DC Signal Linearity in
10mV/A Range (Typical)
Phase versus
frequency, typical
See Appendix 4 Phase versus Frequency
at 1 A Peak (Typical)