Oxygen monitor lcd display identification – Precision Medical Oxygen Monitor User Manual
Page 9
8
O
xygen
M
OnitOr
ITEM NAME
DESCRIPTION
High Alarm Key
Pressing the High Alarm Key when the keypad
is unlocked, the high alarm set point will flash. The
high alarm setting can be adjusted with the
UP/
DOWN keys. When the high alarm value is changed,
pressing the key will save the setting. The Oxygen
Monitor will also save the setting and revert to normal
operation if no keys are pressed within 10 seconds.
The High Alarm Set is adjustable up to 100% Oxygen.
To disable the HIGH Alarm, raise the HIGH Alarm
setting above 100%. Dashes (---) will be displayed next
to the UP key on the LCD display.
NOTE: When the High alarm is disabled dashes (---)
will appear next to the UP key on the LCD display,
the Low alarm will still function.
NOTE: Factory Preset = 50% Oxygen.
High Alarm LED
During a High Alarm condition, the red LED will flash
accompanied by the triple pulse audible alarm.
Up / Down
Keys
The Up/Down Keys are used in conjunction with the
alarm set keys. Pressing either of these keys will
change the alarms set points by 1% increments or 5%
if held down continuously.
Sensor Cable
Connector
Cable Interface connection between Oxygen Monitor
and Oxygen Sensor Cable.
OXYGEN MONITOR LCD DISPLAY IDENTIFICATION
High Alarm Set Point
Measured Gas
High Alarm Indicator
Calibration Failed
Passed
Check Sensor
Low Alarm Indicator
Audio Alarm Paused
Calibration Mode
Mandatory Action
Oxygen Concentration
Low Alarm Set Point
Locked Indicator
Low Battery Indicator