Oxygen monitor lcd display identification – Precision Medical Oxygen Monitor User Manual

Page 9

background image

8

O

xygen

M

OnitOr

ITEM NAME

DESCRIPTION

High Alarm Key

Pressing the High Alarm Key when the keypad
is unlocked, the high alarm set point will flash. The
high alarm setting can be adjusted with the

UP/

DOWN keys. When the high alarm value is changed,
pressing the key will save the setting. The Oxygen
Monitor will also save the setting and revert to normal
operation if no keys are pressed within 10 seconds.
The High Alarm Set is adjustable up to 100% Oxygen.
To disable the HIGH Alarm, raise the HIGH Alarm
setting above 100%. Dashes (---) will be displayed next
to the UP key on the LCD display.
NOTE: When the High alarm is disabled dashes (---)
will appear next to the UP key on the LCD display,
the Low alarm will still function.
NOTE: Factory Preset = 50% Oxygen.

High Alarm LED

During a High Alarm condition, the red LED will flash
accompanied by the triple pulse audible alarm.

Up / Down

Keys

The Up/Down Keys are used in conjunction with the
alarm set keys. Pressing either of these keys will
change the alarms set points by 1% increments or 5%
if held down continuously.

Sensor Cable

Connector

Cable Interface connection between Oxygen Monitor
and Oxygen Sensor Cable.

OXYGEN MONITOR LCD DISPLAY IDENTIFICATION

High Alarm Set Point

Measured Gas
High Alarm Indicator
Calibration Failed

Passed

Check Sensor
Low Alarm Indicator

Audio Alarm Paused

Calibration Mode

Mandatory Action

Oxygen Concentration

Low Alarm Set Point

Locked Indicator

Low Battery Indicator