beautypg.com

Quick setup guide, Sensor installation caution, Warning – Precision Medical Oxygen Monitor User Manual

Page 12

background image

11

O

xygen

M

OnitOr

QUICK SETUP GUIDE

1. Install 4 AA Alkaline Batteries.

2. Connect the Oxygen Sensor and Diverter Fitting.

3. Connect the Coiled Cable.

4. Calibrate the Oxygen Monitor.

5. Set the High/Low Alarms.

SENSOR INSTALLATION

CAUTION

• Inspect the Oxygen Sensor and Diverter Fitting for visual damage or

electrolyte leakage before use. DO NOT USE if damaged.

• Use ONLY an Oxygen Sensor specified by Precision Medical, Inc.

• The Oxygen Sensor should not be used in the presence of flammable

anesthetics such as Diethyl Ether or Cyclopropane

WARNING

DO NOT attempt to open or repair the Oxygen Sensor.

• The Sensor electrolyte is corrosive, and contains lead.

DO NOT let it come in contact with the skin. If it does, flush affected

area with water.

• Check the Sensor regularly for leaks. If the Sensor is leaking, replace

with NEW Sensor. Leaking or used Sensors should be handled and

disposed of in accordance with local regulations.

• An MSDS is available from Precision Medical, Inc.

• If the Oxygen Sensor is used in breathing circuits, the Diverter must be

attached to the Sensor and must be used with the Tee Adapter.

• The Oxygen Sensor must be installed before the Oxygen Monitor can

be operated.

1. Screw the Diverter to the bottom of the Oxygen Sensor, tighten until snug.

2. If using the Tee Adapter, attach to the Diverter.

3. Insert the one end of the Coiled Cable into the top of the Sensor, and

secure by tightening the Locking Ring, until snug.

4. Insert the other end of the Coiled Cable into the Sensor Cable Connection

located on the right side of the Oxygen Monitor. Secure it in place by

tightening the Locking Ring, until snug.

5. Wait approximately 20 minutes for the NEW Sensor to stabilize to the

environment.

6. Calibrate the Oxygen Monitor with the NEW Sensor.