TROUBLESHOOTING

If the oxygen monitor fails to function, consult the Troubleshooting Guide. If

the problem cannot be solved by using Troubleshooting Guide, consult your

Provider.

Problem

Probable Cause

Remedy

Low Battery Indicator

is displayed with a 30

second chirp alarm

1. Battery voltage too

low

1. Replace with 4 NEW AA

alkaline batteries

Check Sensor appears on

display with a continuing

pulsing alarm

1. Cable connections

are not secure

2. Oxygen Sensor not

functioning

3. Using Oxygen Sensor

other than Precision

Medical Oxygen

Sensor

1. Make sure cable

connections are secure

and locking rings are tight

2. Replace with New

Precision Medical

Oxygen Sensor

3. Attach New Precision

Medical Oxygen Sensor

New Oxygen Sensor

responds slowly or seems

to drift

1. Oxygen Sensor has

NOT temperature

stabilized

1. Wait approximately 20

minutes for Oxygen

Sensor to stabilize with

the environment, and

then recalibrate the

Oxygen Monitor

Oxygen Sensor does

not react to changes in

oxygen concentration

1. Condensation on the

Oxygen Sensor

2. Non functioning

Oxygen Sensor

1. Remove Condensation

2. Replace with New

Precision Medical

Oxygen Sensor

Triple Pulse Alarm and

flashing LED

1. Oxygen readings are

outside the High/Low

Alarm limits

2. Loss of Air or Oxygen

Supply

1. Adjust the High/Low Alarm

setting to be above/below

the oxygen value being

displayed

2. Reconnect the Air or

Oxygen Supply

Keys inoperable

(Power ON)

1. The keypad is Locked 1. Unlock the keypad

No Display / LCD screen

will not power ON

1. Dead Batteries

2. Battery installed

incorrectly

1. Check and replace with 4

New AA alkaline batteries

2. Check that batteries are

installed in the proper

position (+/-)

“SERVICE NEED” appears

on the display

1. Electronic malfunction 1. Oxygen Monitor must be

serviced

“CAL FAIL” appears on

the display

1. Improper or wrong

Oxygen Sensor

2. Improper Air / Oxygen

source

3. Non functioning

Oxygen Sensor

1. Recalibrate the Oxygen

Monitor

2. Check the Air / Oxygen

source

3. Replace with New

Precision Medical

Oxygen Sensor