6
O
xygen
M
OnitOr
Low Alarm
Set
High Alarm LED
Low Alarm LED
High Alarm Set
Up
Down
Audio Alarm Paused
On/OFF
Lock / Unlock
Calibrate
Sensor Cable Connector
COMPONENT DESCRIPTION
OXYGEN MONITOR IDENTIFICATION
Low Battery Indication:
Low battery icon
displayed on graphics
screen, and audible alarm
Alarm System:
high/low alarms, respective flashing red LEDs and
graphics, 68db audible alarm @ 1 meter
Low Alarm Range:
*15% - 99% Oxygen
(*Requires extra action to set below 18%)
High Alarm Range:
18% - 100% Oxygen
High Alarm
De-Activation Setting:
Above 100% (“---” will appear )
Alarm Accuracy:
Displayed value +/- 0.1
Patient Contact:
Indirect contact via gas passing through sensor
sampling site.
Sensor Specifications
Sensor Type
Precision Medical 504877 galvanic oxygen sensor
(fuel cell)
Expected Sensor Life
> 1,000,000 O2 % Hours
Specifications are subject to change without notice.