REPLACEMENT PARTS

Description

Part #

User Manual

505127

Precision Medical Oxygen Sensor with Diverter

504877

Tee Adaptor

505126

Extendible Cable

504937

AA Alkaline Battery (4 Pack)

505124-4

Rubber Feet (4)

505122-4

Battery Cover

504909

V Block

505010

Diverter

505344

ACCESSORIES

Description

Part #

Monitor Wall Mount Bracket

505189

Monitor Vertical Pole Mount Clamp

505013

Monitor Horizontal Pole Mount Clamp

505014

Dove tail Bracket

505012

Dove tail Bracket Screw

505712

Orders for replacement parts should include the part number, if available and the
model and serial number of the instrument for which the parts are intended.

ALARM CONDITIONS

Alarm Condition

Alarm Meaning

Corrective Action

High alarm LED flashing

and auditable beep

Measured O2

concentration is higher

than the high alarm set

point.

• Adjust O2 concentration

source to prescribed

dose.

• Set high alarm to

prescribed limit.

Low alarm LED flashing

and auditable beep

Measured O2

concentration is lower

than the low alarm set

point.

• Adjust O2 concentration

source to prescribed

dose.

• Set low alarm to

prescribed limit.

Low Battery symbol on

and O2 displays “- - -“

instead of a value.

The voltage output of

the installed batteries

is low and will require

replacement soon.

• Replace all batteries with

new batteries.

Oxygen Sensor symbol on.

Oxygen Sensor failure

• Replace O2 sensor

• Recalibrate O2 Monitor