Replacement parts, Accessories, Alarm conditions – Precision Medical Oxygen Monitor User Manual
Page 19
18
O
xygen
M
OnitOr
REPLACEMENT PARTS
Description
Part #
User Manual
505127
Precision Medical Oxygen Sensor with Diverter
504877
Tee Adaptor
505126
Extendible Cable
504937
AA Alkaline Battery (4 Pack)
505124-4
Rubber Feet (4)
505122-4
Battery Cover
504909
V Block
505010
Diverter
505344
ACCESSORIES
Description
Part #
Monitor Wall Mount Bracket
505189
Monitor Vertical Pole Mount Clamp
505013
Monitor Horizontal Pole Mount Clamp
505014
Dove tail Bracket
505012
Dove tail Bracket Screw
505712
Orders for replacement parts should include the part number, if available and the
model and serial number of the instrument for which the parts are intended.
ALARM CONDITIONS
Alarm Condition
Alarm Meaning
Corrective Action
High alarm LED flashing
and auditable beep
Measured O2
concentration is higher
than the high alarm set
point.
• Adjust O2 concentration
source to prescribed
dose.
• Set high alarm to
prescribed limit.
Low alarm LED flashing
and auditable beep
Measured O2
concentration is lower
than the low alarm set
point.
• Adjust O2 concentration
source to prescribed
dose.
• Set low alarm to
prescribed limit.
Low Battery symbol on
and O2 displays “- - -“
instead of a value.
The voltage output of
the installed batteries
is low and will require
replacement soon.
• Replace all batteries with
new batteries.
Oxygen Sensor symbol on.
Oxygen Sensor failure
• Replace O2 sensor
• Recalibrate O2 Monitor