Test #’s problem probable cause remedy – Precision Medical Air-Oxygen Blender User Manual

Page 29

background image

- 28 -

Air-Oxygen Blender

SECTION 4: TROUBLESHOOTING

Test #’s

Problem

Probable Cause

Remedy

1

Pressure drop greater

than 2 psi in two

minutes

Leakage from manifold caused by cut or

missing o-ring or due to particulates.

Check ALL manifold connections (inlets, outlets,

plugs, proportioning valve, alarm poppet, etc.) with

oxygen leak detector to find source of leakage; if leak

is found, remove appropriate parts and clean seal

area and o-rings and/or replace appropriate o-ring.

Aluminum Blocks ONLY

Leakage from diaphragm housing

caused by damaged diaphragm or

particulates.

Check diaphragm housings with oxygen leak

detector. If leak is found, replace with new

Engineered Composite Diaphragm Blocks.

Aluminum Blocks ONLY

Leakage from check valve/diaphragm

housing seal caused by cut or missing

o-ring or due to particulates.

Check check-valve/diaphragm housing seal with

oxygen leak detector. If leak is found, replace with new

Engineered Composite Diaphragm Blocks.

Ball not sealing in the alarm bypass.

Replace spring and ball in alarm bypass; ensure

seal surface is clean.

Auxiliary bleed is open.

Close auxiliary bleed by turning and pulling knurled

collar away from cover until bleed is closed.

Leakage from one of the outlets.

Replace outlet.

2 and 3

Back flow leak

Faulty inlet.

Replace duckbill valve or entire inlet assembly.

4 thru 8

Measured

F

IO

2

values

do not meet target

values

Outlet flow is less than 3 lpm.

Adjust flowmeter to 3 lpm (

Note: flow must be

adjusted after each change in

F

IO

2

setting).

Proportioning valve endpoints are not

set correctly.

Set proportioning valve endpoints (See setup

procedure in Section E)

.

Diaphragm blocks not balancing

properly.

Replace with new Engineered Composite

Diaphragm Blocks.

Internal leakage in proportion valve.

Remove proportioning valve; clean seal areas and/

or replace the two rear o-rings. If necessary, replace

proportioning valve assembly.

9 thru 11

Measured

F

IO

2

values

do not meet target

values

Bleed not open.

Open bleed by turning and pushing the knurled

collar until it contacts the cover.

Blockage in bleed holes.

Replace auxiliary outlet.

Internal leak in proportioning valve.

Remove proportioning valve; clean seal areas and/

or replace the two rear o-rings. If necessary, replace

proportioning valve assembly.

Flow not set to 1 lpm.

Adjust flow to 1 lpm.

Diaphragm blocks not balancing properly. Replace diaphragm block.

12 and 14

Alarm not audible and

gas is not exiting the

alarm vent

Pressure differential not sufficient to

trigger alarm.

Ensure supply pressures are set properly to achieve

differential (

Low Flow: 18 to 22; High Flow: 13 to 25).

Alarm not audible and gas

is exiting the alarm vent

Faulty alarm.

Replace alarm.

(See Figure C & M)

13 and 15

Alarm does not turn off

after balancing supply

pressures

Faulty alarm assembly.

Replace alarm assembly.

(See Figure D & K)

Ball not sealing in the alarm bypass.

Replace spring and ball in alarm bypass; ensure

seal surface is clean.

16 thru 19

Measured flow values

do not meet minimum

target values

Gas inlets are restricted.

Check appropriate gas inlet(s) for restriction in

gas pathway; replace duckbill or entire inlet as

necessary.

High flow model only: low flow inlets

installed in place of high-flow inlets.

Confirm that high-flow inlets are installed; replace

as necessary.

Alarm bypass is threaded too far into manifold

block (only applicable to tests 17 and 18).

Replace ball and spring (see Figure I for proper

assembly method).

High flow models only: wrong ball in

alarm bypass block (only applicable to

tests 17 and 18).

Confirm that correct ball is installed in alarm

bypass; replace as necessary.