beautypg.com

High flow, Figure f, Figure g – Precision Medical Air-Oxygen Blender User Manual

Page 20: Figure h

background image

- 19 -

Air-Oxygen Blender

HIGH FLOW

Figure F

16. Remove each diaphragm housing assembly [20]

from the manifold by removing the two hex socket

head screws [21] on the top of each diaphragm

housing with 5/32 hex key.

NOTE: If you have received engineered composite

blocks, keep the four (4) Hex Socket head

screws [21] and discard the aluminum

blocks you have removed.

[20]

QUANTITY - 2

[21]

QUANTITY - 4

Figure G

17. Using 5/32 in. long hex wrench, unscrew and

remove plug [22] from the side of the manifold

block. (For High Flow models with a manifold outlet

assembly remove plug from left side of manifold

block instead of from the bottom of the manifold

block.)

18. Remove rear plug [22] from the back of the manifold

block. Insert a long 5/32 hex key through rear plug

opening, unscrew and remove the alarm bypass

body [23], ball [24] and spring [25].

QUANTITY - 2

[23]

[25]

[24]

[22]

[22]

[22]

NOTE:

For

Engineered Composite Diaphragm Blocks disregard Figure “H” and proceed to Figure “I”

in the Blender maintenance process. Diaphragm blocks shall be installed in pairs and aluminum

blocks shall not be mixed with engineered composite blocks.

Figure H

19. Remove and discard the two O-rings [26] from the

bottom of each of the diaphragm housings [27].

20. Remove the four hex socket head screws [28] holding

diaphragm housings together.

21. Remove and discard the diaphragm [29].

22. Using Retaining Ring Pliers, remove and discard check

valves [30]. Ensure inner O-ring [36] is removed from

diaphragm housing.

NOTE: If check valve [30] comes apart when removing

it from diaphragm housing, a ¼ in. hex wrench

or (#3 Phillips Screwdriver) will be needed to

complete removal of the check valve from the

diaphragm housing.

23. Repeat steps 19-22 for the remaining diaphragm

housing assembly.

24. Make sure all O-rings have been removed from

diaphragm housings and check valves.

25. Disassembly is complete.

26. Manifold block may be ultrasonically cleaned.

Aluminum Blocks

[27]

[26]

QUANTITY - 2

QUANTITY - 4

[28]

QUANTITY -4

[29]

[30]

QUANTITY- 2

[36]

QUANTITY -2