Air-Oxygen Blender

USA and CANADA ONLY

HIGH Flow Operation Verification Procedure

(50 psi / 3.45 bar MODELS)

Serial No.

S
E

Q
#

DIAL

SET

Oxygen%

OXYGEN

PRESS

±1.0

AIR

PRESS

±1.0

FLOWMETER

SET TO

lpm ±0.2

AUXILIARY

BLEED

FUNCTION

TARGET

VALUE

ACTUAL

VALUE

psi

bar

psi

bar

1

ANY

50

3.45

50

3.45

closed

closed

leak

<2 psi

/

2 MIN

*2

60

75

5.17

0

0

0

closed

back flow <100 ml/min

*3

60

0

0

75

5.17

0

closed

back flow <100 ml/min

4

21

50

3.45

50

3.45

15

closed

end point

(±0.3)

Source Value

5

40

50

3.45

50

3.45

15

closed

set point 37.0%-43.0%

6

60

50

3.45

50

3.45

15

closed

set point 57.0%-63.0%

7

80

50

3.45

50

3.45

15

closed

set point 77.0%-83.0%

8

100

50

3.45

50

3.45

15

closed

end point

(±0.3)

Source Value

9

60

50

3.45

50

3.45

1.5

open

set point 57.0%-63.0%

10

60

60

4.14

67

4.62

1.5

open

set point 57.0%-63.0%

11

60

60

4.14

50

3.45

1.5

open

set point 57.0%-63.0%

12

60

50

3.45

Slowly

reduce

to 31

Slowly

reduce

to 2.14

15

closed

Alarm

ON

31.0

± 6.0 psi

13

60

50

3.45

Slowly

Increase

until alarm

shuts off

15

closed

Alarm

OFF

45.0 psi

MAX

14

60

Slowly

Reduce

to 31

Slowly

Reduce

to 2.14

50

3.45

15

closed

Alarm

ON

31.0

± 6.0 psi

15

60

Slowly

Increase

until alarm

shuts off

50

3.45

15

closed

Alarm

OFF

45.0 psi

MAX

16

60

50

3.45

50

3.45

MAX

closed

flow rate

120.0 lpm

MIN

17

60

50

3.45

0

MAX

closed

flow rate

85.0 lpm

MIN

18

60

0

50

3.45

MAX

closed

flow rate

85.0 lpm

MIN

19

60

50

3.45

50

3.45

MAX

open

flow rate

120.0 lpm

MIN

* Reference, Letter D. (Reverse Gas Flow Procedure)