background image

- 11 -

Air-Oxygen Blender

LOW FLOW

NOTE: See

Figure J-1 for Aluminum Diaphragm Blocks and Figure J-2 for Engineered

Composite Diaphragm Blocks. Diaphragm blocks shall be installed in pairs and

Aluminum Blocks shall not be mixed with Engineered Composite Blocks.

Figure J-1

(Aluminum Diaphragm Blocks) (shown)

5. Assembling diaphragm housing assemblies.

a. Ensure all O-rings have been removed from diaphragm

housing.

b. Place diaphragm alignment tool through center hole of one

diaphragm housing.

c. Place new diaphragm [29] on top of diaphragm housing, ensure

center pin of diaphragm is in the center of diaphragm alignment

tool (P/N 504838).

d. Place other diaphragm housing on top of the one with the

diaphragm. Ensure the inlet holes of the diaphragm housings

are on the same side.

e. Fasten the two diaphragm housings together using four hex socket

head screws [28]. Torque screws to 60 in-lbs.

f. Place four new O-rings [26] on the inlet holes of the diaphragm

housings.

g. Remove diaphragm alignment tool.

h. Place new O-rings [36] onto each check valve [30].

i. By hand, carefully insert new check valve [30] into diaphragm

housing. (Check valves are double threaded) Thread check valves

into diaphragm housing using Retaining Ring Pliers. Check valve

must be flush with surface of diaphragm housing. DO NOT USE

power tool to tighten. Repeat this step for the second check valve.

j. Using two long hex socket head screws [21] secure each

diaphragm housing assembly to the manifold block. Torque to

60 in-lbs (6.8 Nm).

k. Repeat steps a - i for remaining diaphragm housing assembly.

Figure J-2

(Engineered Composite Diaphragm Blocks)

5. Installing diaphragm housing to manifold.

a. Using two long hex socket head screws [21] secure each

diaphragm housing assembly to the manifold block. Torque to

30 in-lbs (3.4 Nm).

[28]

QUANTITY - 4

[29]

[26]

QUANTITY - 4

[30]

QUANTITY - 2

[21]

QUANTITY - 2

[36]

QUANTITY - 2

Diaphragm

Alignment

Tool

Aluminum Blocks

Figure K

6. Lubricate alarm bypass assembly bore on manifold block

with Krytox GPL 106.

7. Using retaining ring pliers install retaining ring [12]

in one side of alarm assembly bore.

8. Insert alarm assembly [13] into bore of manifold block.

9. Install remaining retaining ring.

10. For Models without manifold block assembly:

Replace O-ring [32] on plug [22] then install plug [22] into

bottom hole of manifold block.

11. Thread auxiliary outlet assembly [10] to manifold block,

torque to 10 ft-lbs (13.6 Nm).

[12]

QUANTITY - 2

[10]

[13]

[32]

[22]