- 10 -

Air-Oxygen Blender

Step 2: CLEANING

Precision Medical, Inc. recommends using an ultrasonic cleaner for cleaning all

non-elastometric and non-metallic components. However, cleaning with an all-purpose

liquid cleaner and rinsing with clean, warm water may be substituted. Both methods require

thoroughly blow drying all passages before reassembly. Follow the ultrasonic cleaner

manufacturer instruction.

Step 3: LOW FLOW ASSEMBLY

Tools Required

Lint Free Swab (optional)

Pointed instrument for removing O-rings

Krytox GPL 106 or equivalent

Oxygen safe lubricant

#2 Phillips Screwdriver

1/2 in. Open End Torque Wrench

(Torque wrench(s) capable of 60 in-lbs and 10 ft-lbs)

11/32 in. Nut Driver

5/32 in. Long Hex Key

Small Retaining Ring Pliers

Tools Supplied with Kit

Diaphragm Alignment Tool

(P/N 504838)

NOTE: For use with Aluminum

Diaphragm Blocks ONLY.

Figure I

1. Position manifold block so that the mounting

bracket is facing up. The large holes opposite
the mounting locations of the diaphragm housing
assemblies are facing you.

2. Cover the alarm assembly thread location with

thumb. Drop spring [25] then ball [24] into alarm
bypass orifice.

3. Place alarm bypass body [23] onto long shaft

of 5/32 hex key. Guide the alarm bypass body,
threads first into the alarm bypass orifice, screw into
cavity. Tighten until alarm bypass body is flush with
bottom of center hole. Tighten enough to clear
center threads ONLY.

4. Replace O-ring [32] on plug [22]. Insert plug and

tighten.

[32]

[23]

[25]

[24]

[22]