Air-Oxygen Blender
Step 2: CLEANING
Precision Medical, Inc. recommends using an ultrasonic cleaner for cleaning all
non-elastometric and non-metallic components. However, cleaning with an all-purpose
liquid cleaner and rinsing with clean, warm water may be substituted. Both methods require
thoroughly blow drying all passages before reassembly. Follow the ultrasonic cleaner
manufacturer instruction.
Step 3: LOW FLOW ASSEMBLY
Tools Required
Lint Free Swab (optional)
Pointed instrument for removing O-rings
Krytox GPL 106 or equivalent
Oxygen safe lubricant
#2 Phillips Screwdriver
1/2 in. Open End Torque Wrench
(Torque wrench(s) capable of 60 in-lbs and 10 ft-lbs)
11/32 in. Nut Driver
5/32 in. Long Hex Key
Small Retaining Ring Pliers
Tools Supplied with Kit
Diaphragm Alignment Tool
(P/N 504838)
NOTE: For use with Aluminum
Diaphragm Blocks ONLY.
Figure I
1. Position manifold block so that the mounting
bracket is facing up. The large holes opposite
the mounting locations of the diaphragm housing
assemblies are facing you.
2. Cover the alarm assembly thread location with
thumb. Drop spring [25] then ball [24] into alarm
bypass orifice.
3. Place alarm bypass body [23] onto long shaft
of 5/32 hex key. Guide the alarm bypass body,
threads first into the alarm bypass orifice, screw into
cavity. Tighten until alarm bypass body is flush with
bottom of center hole. Tighten enough to clear
center threads ONLY.
4. Replace O-ring [32] on plug [22]. Insert plug and
tighten.
[32]
[23]
[25]
[24]
[22]