- 22 -

Air-Oxygen Blender

HIGH FLOW

Figure L

11. Replace O-ring [32] on plug [22]. Install plug and

tighten.

[22]

[32]

Figure M

12. Start threads of new alarm assembly [11] by hand,

tighten with Retaining Ring Pliers, ensure not to bend

reed.

13. Attach bottom cover [6] using four flat head screws [2].

14. Install new air inlet assembly [4], torque to 10 ft-lbs

(13.6 Nm).

15. Install new oxygen inlet assembly [5] torque to

10 ft-lbs (13.6 Nm). Oxygen inlet assembly has left

handed threads.

16. Thread new primary outlet assembly [37] to manifold

block torque to 10 ft-lbs (13.6 Nm).

[6]

[2]

QUANTITY- 4

[4]

[5]

[11]

[37]

Figure N

17. Lubricate proportioning valve bore with Krytox GPL 106.

18. Align the (3) holes on the proportioning valve assembly

[16] equal distance between the (2) plugs and push in.

*Reference drawing below.

19. Replace washer [19] and phillips head screw [18].

20. Place new resistance ring [17] in its place on the

proportioning valve assembly.

*Proportional valve assembly inserted, with the

“H” stamped on back for High Flow Model.

plugs

[16]

[18]

[19]

[17]

The “H” is stamped on the
back for HIGH Flow models