beautypg.com

Precision Medical Air-Oxygen Blender User Manual

Page 18

background image

- 17 -

Air-Oxygen Blender

II. HIGH FLOW (PM5200 Model)

Step 1: DISASSEMBLY

Tools Required

#2 Phillips Screwdriver

#3 Phillips Screwdriver
1/2 in. Open End Wrench

11/32 in. Nut Driver
5/32 in. Long Hex Key
Small Retaining Ring pliers

NOTE: Photos shown of Diaphragm Blocks are Engineered Composite Blocks.

Figure A

1. Rotate dial [1] to the 60 graduation.

2. Remove the two flat head screws [2] on each

side of the top cover [3].

3. Remove top cover by pulling upwards.

The cover will not come off unless the dial

is at the 60 graduation.

[1]

[2]

QUANTITY - 4

[3]

Figure B

4. Use a ½ in. open end wrench to unscrew

and remove the air [4] and oxygen [5] inlet

assemblies and primary [9] outlet assemblies

from bottom of the Blender.

Oxygen inlet has left hand threads.

5. If manifold outlet assembly is present, unscrew

the stem using a 5/32 hex key and holding the

manifold outlet assembly to the bottom of the

Blender.

6. Remove the four flat head screws [2] from

bottom cover [6].

7. Remove bottom cover.

[5]

[4]

[2]

QUANTITY - 4

[6]

[9]