Step 4: test, Typical test configuration diagram – Precision Medical Air-Oxygen Blender User Manual
Page 14

- 13 -
Air-Oxygen Blender
Step 4: TEST
Equipment Required
Medical Air Supply
Medical Oxygen Supply
Calibrated Oxygen Analyzer/Monitor
Calibrated Air or Oxygen flow monitor 0 to 70 lpm or greater
Flowmeter 0 to 70 lpm or greater
Calibrated Pressure Gauges 0 to 100 psi
Regulators Tubing / Hoses
Tools Required
Phillips Screwdriver
Torque Driver capable of 10 in-lbs
Adjustable Wrench
Retaining Ring Pliers
Nut Driver
Typical Test Configuration Diagram
Blender
Calibrated Flow
Monitor
Compressed
Air Supply
Adjustable
0-100 psi
(0-6.89 bar)
Pressure Gauge
0-100 psi
(0-6.89 bar)
Pressure Gauge
0-100 psi
(0-6.89 bar)
Oxygen
Supply
Adjustable
0-100 psi
(0-6.89 bar)
Oxygen Analyzer
Oxygen Flowmeter
0-70 lpm
Condensation Trap
(Recommended)
T - Fitting
INSTRUCTIONS FOR TESTING
A. Air and Oxygen Supply Setup
Both gas supplies must be clean and dry per the specifications outlined in this manual
and have the ability to generate up to 100 psi (6.89 bar) for both the air and oxygen
inlet pressures. Verify air and oxygen supply concentrations with a analyzer/monitor.