Precision Digital PD6060 User Manual
Page 69
Model PD6060 Dual-Input Process Meter
Instruction Manual
69
Display Parameter Action/Setting
Calib
Calibrate
Calibrate 4-20 mA output (internal reference
source used for scaling the output)
4 nmA
4 mA output
Enter mA output value read by milliamp
meter with at least 0.001 mA resolution
20 nmA
20 mA output
Enter mA output value read by milliamp
meter with at least 0.001 mA resolution
User
User I/O
Assign function keys and digital I/O
F1
F1 function key
Assign F1 function key
F2
F2 function key
Assign F2 function key
F3
F3 function key
Assign F3 function key
F4
F4 function key
Assign F4 function key (digital input)
dI I
Digital input 1
Assign digital input 1 – 8, if expansion
modules are connected
dO 1
Digital output 1
Assign digital output 1 – 8, if expansion
modules are connected
ICal
Internal
calibration
Enter internal calibration (used for
recalibrating the meter with a calibrated
signal source)
Ch-A
Channel A
Perform calibration on channel A
Ch-b
Channel B
Perform calibration on channel B
C CAL
Current
calibration
Calibrate 4-20 mA current input (internal
reference source used for scaling the input)
C lo
Current low
Calibrate low current input (e.g. 4 mA)
C Hi
Current high
Calibrate high current input (e.g. 20 mA)
v CAL
Voltage
calibration
Calibrate voltage input
v Lo
Voltage low
Calibrate low voltage input (e.g. 0 V)
v Hi
Voltage high
Calibrate high voltage input (e.g. 10 V)
Copy
Copy
Enter copy function
Send
Send
Send meter settings to another meter
Done
Done
Copy function completed
Diag
Diagnostics
Display parameter settings