beautypg.com

Flow meter control, Calibration and configuration of the flow meter – Electrolux FOM 71 CLS User Manual

Page 155

background image

3

106

104

102

100

98

96

0,5 1 1,5 2 2,5 3 3,5

Series 1

P (bar)

V

olume %

Calibration and configuration of the flow meter

Configure the flow meter according to the
attached configuration scheme. It is very easy to
make mistakes during the configuration so be
careful and follow strictly the scheme. For a full
description of all possibilities see the attached
manual for the Batch controller type 8025. The
calibration factor ”K-factor” must be calibrated
according to conditions in each individual
laboratory. The K-factor depends among other
things of available water pressure. The figure
shows the dependency as measured in the
Electrolux- Wascator laboratory.

The two programmable volumes A1 and A2 shall
be programmed according to the water volume
indicated in EN 60 456. Note that carry over can
depend on textile quality and always make a
check before any new test series.

Flow meter control

6

Fig

5

This manual is related to the following products: