Chapter 3: Commands and Programming

DATA RANGE SETUP Examples

{5,1,2}

5 = DATA RANGE SETUP
1 = Transfer data from CH1
2 = Filtered

d

2

/dt

2

data (when filtering is turned on

with Command 1)

{5,2,0,256,300}

5 = DATA RANGE SETUP
2 = Transfer data from CH2
0 = Filtered raw data (when filtering is turned on
with Command 1)
256 = Starting sample number
300 = Ending sample number

Command 6 - MULTIMETER MODE SETUP

Command 6 specifies a channel for multimetering. Only one channel (the
one you specify) is enabled in the Multimeter Mode at a time.

Syntax

{6, Set/Reset, Channel Select, Operation}

• Pressing the [MODE] key and entering the Multimeter Mode makes it

possible to use the data analyzer as a stand-along measuring instru-
ment without commands from a graphic scientific calculator.

Set/Reset Parameter

This parameter executes a Multimeter Mode set or reset operation. The initial
default for this parameter is 1.

0

Returns to the Communications Mode

1

Enters the Multimeter Mode

The following points apply whenever the data analyzer is in the Multimeter
Mode.

• Command 6 with a set/reset parameter of 0 is the only valid command

accepted by the data analyzer.

• The default sample time is approximately 1 second.

• Sampling results appear on the display, but are not sent to a connected

calculator.

Channel Select Parameter

This parameter selects a channel for measurement.

1

Channel 1 (CH1)

2

Channel 2 (CH2)

3

Channel 3 (CH3)