Chapter 3: commands and programming – Casio EA-100 Commands and Programming User Manual
Page 16

38
Chapter 3: Commands and Programming
1
Auto-ID
If Auto-ID is not activated, voltage in the range 0 to 5V is used.
2
Voltage (
±
10V)
3
Current (
±
10A)
Based on Vin input, 1V = 1A.
4
Resistance (1k
Ω
to 100k
Ω
)
5
Period (0.000004 to 50 seconds)
See “Period and Frequency Measurement” in Appendix A.
6
Frequency (0.02Hz to 250kHz)
See “Period and Frequency Measurement” in Appendix A.
7
Temperature (–20
°
C to 130
°
C)
8
Temperature (–4
°
F to 266
°
F)
9
Light (100 to 999)
10
Voltage (0 to 5V)
Post-Processing Parameter
This parameter specifies the type of post-processing to be applied to sam-
pled data. The initial default for this parameter is 0.
0
None (no processing)
1
d/dt
(first derivative)
Only the
d/dt
indicator is on the display when this type of post-
processing is selected.
2
d/dt
and
d
2
/dt
2
(first derivative and second derivative)
The
d/dt
and
d
2
/dt
2
indicators are on the display when this type
of post-processing is selected.
3
Statistics (mean, standard deviation, minimum, maximum)
Derivative and statistical post-processing are performed using values pro-
duced by conversion equations.
• Post-processing is disabled whenever the number of samples parameter
of Command 3 (SAMPLING/TRIGGER SETUP) is –1, which specifies
real-time display of sampled data.
Stat Samples Parameter
This parameter is used during statistical post-processing to specify how many
samples should be used for post-processing. This parameter is valid only
when 3 (statistics) is specified for the post-processing parameter. A value
from 2 to 512 can be specified for this parameter, and The initial default for
this parameter is 10.
P.68
P.68