Ion Science GasCheck IS User Manual
Page 36
GASCHECK 5000IS MANUAL
Ion Science Ltd
Page 36 of 40
Unrivalled Detection. www.ionscience.com
Maintenance
Fault finding
Fault symptom
Diagnosis
Action
Unit will not switch on
No/dead batteries fitted
Try another battery set (safely)
Unit will not switch off
Software locked out
Remove & replace battery clip
(safely)
Battery symbol appears
Battery state low
Replace batteries (safely)
Back light drains battery
life
Excessive use of backlight
Avoid or reduce use of backlight
Unit will not zero
Software locked out by
hardware
Remove & replace batteries (safely)
LCD appears blank/dim
Contrast control setting wrong
Turn off & on; press & hold contrast
Bar graph/number do not
change
Peak hold selected
Press zero or change Peak Hold
state
Range does not change
Range not set in Auto-range
Press & hold Range until Auto-range
He range will not show –5 Unit with nozzle ON and/or
range
Reset via Nozzle and Range keys
Gas group is wrong
Gas Group not set correctly
Press & hold Gases until shown OK
Nozzle/Peak not showing
Nozzle/Peak is not set
correctly
Press & hold Nozzle until both
correct
Calibration due message
Unit has exceeded 1 year on
cal
Return to Ion Science for
recalibration
Unit will not zero correctly Sensor contaminated
Leave to stabilise
Blocked probe
Probe blocked by solid or
liquid
Remove probe; clean; check now OK
Stalled motor
Motor failed & probe not
blocked
Return to supplier for repair
Instrument readings noisy Loose probe
Tighten nut slightly
Instrument readings noisy Test environment
contaminated
Move test into clean area
5000is will not switch on
Internal fuse blown
Return to supplier for repair
Table 5 – Fault finding