beautypg.com

Rear view, Installation, Power supply and connection – HEIDENHAIN ND 261 v.2 User Manual

Page 10: Ambient conditions

background image

Rear View

RS-232-C/V.24
data interface

Power switch

Installation

You can fix the display unit to a flat surface with M4
bolts (see illustration at right).

The units can also be stacked. Adhesive inserts (in-
cluded in delivery) prevent them from sliding.

172

±

0.2

6.77

±

.008"

140

±

0.2

5.51

±

.008"

Interfaces X1, X31 and X41 interfaces comply with the recommendations
in EN 50178 for separation from line power.

ND 261 V, Id. Nr. 322 352 ..
Angle encoder input: 1 V

PP

Connecting cable: max. 60 m
Max. input frequency: 300 kHz

ND 261 A, Id. Nr. 283 482 ..
Angle encoder input: 11 µA

PP

Connecting cable: max. 30 m
Max. input frequency: 100 kHz

Ground terminal

X1

X31 (V.24 /RS-232-C)

X41 (EXT)

100 ... 240 V

50 ... 60 Hz

D-sub connection
EXT

Danger of electrical shock!
Unplug the power cord before opening the housing.
Connect a protective ground. This connection must never be interrupted.

Danger to internal components!
Do not engage or disengage any connections while the unit is under power.
Use only original replacement fuses.

To increase noise immunity, connect the ground terminal on the rear panel to
the central ground point of the machine. (Minimum cross section of the con-
necting cable: 6 mm²).

Power Supply and Connection

Primary-clocked power supply.
Voltage range: 100 V to 240 V (– 15% to + 10%) Frequency: 48 Hz to 62 Hz
Power consumption typ. 8 W Line fuse: F 1 A (in unit)
Minimum cross-section of the power line: 0.75 mm

2

Ambient Conditions

Temperature range

Operation: 0 °C to + 45 °C (32 °F to 113 °F)
Storage: – 30 °C to + 70 °C (– 22 °F to 158 °F)

Rel. humidity

Annual average: < 75%; maximum: < 90%

Weight

1.5 kg

284 250-26 · SW 246 183 01 · 1 · 2/2000 · H · Printed in Germany · Subject to change without notice

DR. JOHANNES HEIDENHAIN GmbH
Dr.-Johannes-Heidenhain-Straße 5
D-83301 Traunreut, Germany
(0 86 69) 31-0 . 56 832
(0 86 69) 50 61

FAX

FAX

Service (0 86 69) 31-12 72

TNC-Service (0 86 69) 31-14 46
(0 86 69) 98 99

Tx

http://www.heidenhain.de

This manual is related to the following products: