Switching Inputs and Outputs EXT (X41)

Inputs at D-sub connection EXT (X41)

Pin

Function

1, 10

0 V

2

Reset display to zero, clear error message

3

Set display to the value selected in P79

4

Ignore reference mark signals

5

Start series of measurements

1)

6

Externally select display value for series of
measurements

1)

7

Display MIN value of series of measurements

1)

8

Display MAX value of series of measurements

1)

9

Display difference MAX – MIN

1)

22

Pulse: Output measured value

23

Contact: Output measured value

25

Enable or disable REF mode
(current REF status is changed)

12, 13, 24

Do not assign

11, 20, 21

Vacant

Special case: Display current measured value ACTL

If you wish to display the current measured value ACTL of a
series of measurements, note for inputs 7, 8 and 9:
Either none or more than one of these inputs must be active.

Switching Inputs and Outputs EXT (X41)

Danger to internal components!
Voltage sources for external circuitry must conform
to the recommendations in EN 50 178 for low-
voltage electrical separation. Connect inductive
loads only with a quenching diode parallel to the
inductance.

Only use shielded cable!
Connect the shield to the connector housing.

Outputs at D-sub connection EXT (X41)

Pin

Function

14

Display value is zero

15

Measured value

trigger limit A1 (P62)

16

Measured value

trigger limit A2 (P63)

17

Measured value < lower sorting limit (P18)

18

Measured value > upper sorting limit (P19)

19

Error

(see “Error Messages”)

1)

Only in the linear measurement mode