HEIDENHAIN ND 281 B v.1 User Manual
Page 42

42
RS-232-C/V.24 Data Interface (X31)
Measured value output after “STX” signal (Ctrl B)
If the display unit receives the control character STX (Ctrl B)
over the RS-232-C/V.24 interface (X31), it outputs the current
measured value over the interface.
➤
➤
➤
➤
➤ Transfer the control character Ctrl B over the RXD line of
the RS-232-C/V.24 interface (X31).
Characteristic times for measured value output
Process
Time
Storage delay
t
1
≤
1 ms
Measured value output after
t
2
≤
50 ms
Regeneration time
t
3
≥
0
These times are prolonged if functions are active
(for example, series of measurements with DIFF
value display).
Duration of measured value transfer
t
D
=
[s]
Propagation times for measured value output after “Ctrl B”
BASIC program for measured value output with “Ctrl B”
10
L%=18
20
CLS
30
PRINT "V.24/RS-232-C"
40
OPEN "COM1:9600,E,7" AS#1
50
PRINT #1, CHR$ (2);
60
IF INKEY$<>""THEN 130
70
C%=LOC(1)
80
IF C%
90
X$=INPUT$(L%,#1)
100 LOCATE 9,1
110 PRINT X$;
120 GOTO 50
130 END
t
1
t
2
Ctrl B
t
D
t
3
Ctrl B
187 + (11 • number of blank lines)
baud rate
- TNC 122 User Manual (63 pages)
- TNC 122 Technical Manual (70 pages)
- TNC 360 Service Manual (157 pages)
- TNC 416 Technical Manual (510 pages)
- TNC 335 Technical Manual (581 pages)
- TNC 360 User Manual (237 pages)
- TNC 360 ISO-Programmierung (2 pages)
- TNC 415 (280 540) User Manual (227 pages)
- TNC 370D (92 pages)
- TNC 416 (289 pages)
- TNC 415 (280 540) Technical Manual (752 pages)
- TNC 415 (259 96x) Service Manual (195 pages)
- TNC 407 (280 580) User Manual (376 pages)
- iTNC 530 (340 420) Pilot (104 pages)
- TNC 407 (280 580) ISO Programming (333 pages)
- TNC 415 (280 540) Service Manual (252 pages)
- PT 880 Installation (112 pages)
- ND 100 User Manual (116 pages)
- ND 287 User Manual (147 pages)
- ND 280 Quick Start (12 pages)
- ND 200 (156 pages)
- ND 282 (10 pages)
- ND 287 Quick Start (26 pages)
- ND 282 B (39 pages)
- ND 281 A (44 pages)
- ND 281 B v.2 (65 pages)
- ND 221 v.2 (10 pages)
- ND 231 B v.2 (56 pages)
- ND 231 B v.1 (44 pages)
- ND 221 B v.2 (45 pages)
- ND 550 v.2 (8 pages)
- NDP 560 (10 pages)
- ND 523 (93 pages)
- ND 570 (8 pages)
- ND 750 v.2 (46 pages)
- ND 760 v.3 (72 pages)
- ND 770 v.1 (40 pages)
- ND 770 v.3 (41 pages)
- ND 760 E (44 pages)
- IOB 49 (21 pages)
- NDP 960 (68 pages)
- ND 780 Installation (132 pages)
- ND 970 (47 pages)
- ND 1100 Quick Start (36 pages)