40

Operating parameters for measured value output

Parameter

Function

P50 RS232

Baud rate

P51 RS232

Number of additional blank lines for
measured value output

Display freeze during measured value output

In operating parameter P23, you can specify how the
measured value output signal will affect the display unit.

Display freeze during measured value output

P23

Concurrent display, no display freeze: The
display value is the current measured value

DISPL. ACTL.

Frozen display: Display is stopped
(frozen) and updated by every
measured value output signal

DISPL. HOLD

Frozen/concurrent display: Display is
frozen as long as a measured value output
signal is present

DISPL. STOP

RS-232-C/V.24 Data Interface (X31)

187 + (11 • number of blank lines)

Measured value output via PRINT function

In the linear measurement mode
Press MOD repeatedly until the indicator PRINT blinks.
Start the measured value output with ENT.
In the angular measurement mode
Press MOD (this feature can be disabled with operating
parameter 86).

Duration of measured value transfer

t

D

=

[ s]

Indicator preselection (linear measurement mode)

Operating parameter P86 allows you to define which indicator
is displayed first when MOD is pressed.

baud rate