List of operating parameters – HEIDENHAIN ND 281 B v.1 User Manual
Page 26

26
Parameter
Settings / Function
P80 ENT-CL
Set display
No set/Set with
CL/ENT
CL-ENT OFF
Zero reset with CL
setting disabled with ENT
CL......ON
Zero reset with CL and set with
ENT to value selected in P79
CL-ENT ON
P82 DISPL.ON
Message after switch-on
ENT...CL message
MESSAGE ON
No message
MESSG. OFF
P85 EXT.REF
External REF
REF over D-sub port EXT
EXT.REF ON
No REF over D-sub port EXT
EXT.REF OFF
P86 MOD
In the linear measurement mode
First indicator after pressing MOD
START
MIN ACTL MAX DIFF
In the angular measurement mode
PRINT via MOD disabled
PRINT OFF
PRINT via MOD enabled
PRINT ON
Parameter
Settings / Function
P44 REF
Reference mark evaluation
Evaluate the reference mark
REF. ON
Do not evaluate the reference
mark
REF. OFF
P45 ALARM
Encoder monitoring
No monitoring
ALARM OFF
Monitor the frequency
FREQUENCY
Monitor contamination
CONTAMINAT.
Contamination + frequency
FRQ. + CONT.
P50 RS232
Baud rate
110 / 150 / 300 / 600 / 1200 /
2400 / 4800 / 9600 / 19 200 /
38 400 baud
P51 RS232
Additional blank lines
during data output
BK LINE 1
0
≤
P51
≤
99
Default setting: 1
P62 A1
Trigger limit 1
P63 A2
Trigger limit 2
P79 PRESET
Value for datum point
Enter numerical value for
datum setting over switching input
or with ENT key
List of Operating Parameters