List of operating parameters – HEIDENHAIN ND 281 B v.1 User Manual
Page 24

24
Parameter
Settings / Function
P00 CODE
Enter a code number:
9 51 48:
To change the protected
operating parameters
41 52 63: To select the linear or angular
measurement mode
10 52 96: Nonlinear axis error
compensation
24 65 84: To lock the keyboard
66 55 44: To show the software version
24 65 82: Distance-to-go display
P01
Units of measurement
1)
Display in millimeters
MM
Display in inches
INCH
P02 X1/X2
Select an encoder input
Signals at X1
11 µAPP
Signals at X2
1 VPP
P08 DISPL.
Display mode
2)
Decimal Degrees
DEC. DEGREE
Deg., minutes, seconds
DEG.MIN.SEC.
P09 ANGLE
Angle display
2)
+/- 180°
+/- 180 DEG.
+/-
∞
ENDLESS
360°
360 DEG
.
1)
Only in linear measurement mode.
2)
Only in angular measurement mode.
List of Operating Parameters
List of Operating Parameters
Parameter
Settings / Function
P11 SCL
Scaling factor
1)
Scaling factor on
SCALING ON
Scaling factor off
SCALNG. OFF
P12 SCL
Scaling factor
1)
Enter a numerical value
0.100000 < P12 < 9.999999
Default setting: 1.000000
P17 SORT
Sorting and tolerance checking
Sorting and tol. checking ON
SORT ON
Sorting and tol. checking OFF
SORT OFF
P18 L.LIMIT
Lower limit for sorting
P19 U.LIMIT
Upper limit for sorting
P21 SERIES
Display for series of measurements
1)
OFF
MIN MAX ACTL DIFF
P23 DISPL.
Display stop for measured value output
Concurrent display, no display freeze;
the display value is the current actual
value
DISPL. ACTL.
Frozen display; hold display until next
measured value output
DISPL. HOLD
Frozen/concurrent display; freeze display
as long as Pulse/Contact for measured
value output is present
DISPL. STOP